Mass spectrometer

Abstract

The quadrupole Mass Spectrometer of the Induced Environment Contamination Monitor (IECM) operates in the range from 2 to 150 amu. It is pointed out that the Mass Spectrometer on STS-2 performed very well. It was found that the column density of H2O effluent from the Shuttle reached a maximum of 1 x 10 to the 13th per sq cm at 7 hr, 30 min and decreased by a factor of 7.5 during the subsequent 40 hrs. The count rate response of H2O could be correlated with mission-related events, taking into account the dumping of supply water, the operation of the Flash Evaporator System, and the firing of a primary reaction control system engine.