NASA NTRS ยท 19830011810
SUMS calibration test report
Abstract
Calibration was performed on the shuttle upper atmosphere mass spectrometer (SUMS). The results of the calibration and the as run test procedures are presented. The output data is described, and engineering data conversion factors, tables and curves, and calibration on instrument gauges are included. Static calibration results which include: instrument sensitive versus external pressure for N2 and O2, data from each scan of calibration, data plots from N2 and O2, and sensitivity of SUMS at inlet for N2 and O2, and ratios of 14/28 for nitrogen and 16/32 for oxygen are given.
Keep this discovery
Explore connections, maps & timelines
Robertson, G.. 1982-03-01. SUMS calibration test report. https://ntrs.nasa.gov/citations/19830011810
Cite the original work for its findings. Save a collection to share your selection of sources.