Engineering Papers⌕ Search

NASA NTRS · 19810057742

Molecular beam facility for studying mass spectrometer performance

Abstract

An apparatus which produces neutral gas beams in the velocity range from thermal to 6 km/s makes possible studies simulating the motion of instruments through tenuous atmospheres. Use of the apparatus in studying the response of an open-source mass spectrometer is examined. Experiments performed include responses when the instrument is in a normal mode of operation, when the ion source potentials are adjusted to reject striking gas particles, when the beam is restricted so that portions of the ion source are struck, and when the beam strikes at angles other than normal incidence. Results indicate specular radiation of particles out of the source and stagnation ratios close to unity for normal operation. Retarding potential studies dropped the sensitivity by a factor of about 10, and as angles of attack are varied, the effect was found to depend upon beam velocity and the way the ions are initially accelerated.

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Ballenthin, J. O., Nier, A. O.. 1981-07-01. Molecular beam facility for studying mass spectrometer performance. https://ntrs.nasa.gov/citations/19810057742