Abstract

The noise characteristics of thermistor bolometers immersed in layers of arsenic/selenium glass uniform in composition were examined. Using a controlled deposition technique, layers of glass were deposited, thermistor bolometers immersed, and their electrical characteristics measured after various thermal treatments. Markedly improved stability of the detector noise was observed using this new technique.

Keep this discovery

Explore connections, maps & timelines

BibTeXRIS

Rolls, W. H.. 1979-06-16. Selenium immersed thermistor bolometer study. https://ntrs.nasa.gov/citations/19800023269