It was demonstrated that the thickness of the stimulator titanium enclosure is directly related to the battery recharge time cycle. Reduction of the titanium enclosure thickness from approximately 0.37 mm (0.015 inch) to 0.05 mm (0.002 inch) significantly reduced the recharge time cycle and thereby patient inconvenience. However, fabrication of titanium enclosures from the thinner material introduced problems in forming, holding, and welding that required improvement in state of the art shop practices. The procedures that were utilized to resolve these fabrication problems are described.