A subcutaneous channeling probe for implanting long leads
The channeling probe described in the present paper was designed to overcome surgical problems of the type that were encountered when a multichannel radio transmitter had to be implanted in a cow. The probe was made of a flexible but sufficiently stiff 9.5-mm-diam nylon rod, consisting of 46-cm sections for convenience in sterilization and surgical handling. Stainless steel sleaves reinforced the threaded connecting joints. At one end, arrowhead-shaped channeling heads could be attached to produce wide channels for large sensors. The other end was tapered for narrow channels. Postoperative problems were not encountered in the use of this probe in cows, sheep, and dogs.