X-ray single Photon Spectrometers
The research at Yale on superconducting tunnel junctions (STJs) focussed on x-ray detectors, using a 5.9 keV x-ray source. All the Yale tests were carried out at 0.21 K, in a two-stage helium 3 dewar. Careful attention was given to grounding and shielding. All current waveforms were digitially recorded and analyzed subsequently on computer. The imaging device geometry was studied. This allows imaging along the length of the tantalum (Ta) absorber, which is 600 nm thick to provide good x-ray absorption. The quasiparticles produced by single photon absorption rapidly cool to the Ta gap, E(sub g_Ta) = 1.4 meV, and then diffuse. They are trapped in the A1 traps at the two ends (E(sub g_A1) = 0.4 meV), to give two charges that tunnel out of the A1 traps, Q(sub 1) and Q(sub 2). The total charge is Q = Q(sub 1) + Q(sub 2). This lateral trapping is what allows imaging.