Engineering Papers⌕ Search

Engineering topics

LePort, F.

Publications and source records attributed to LePort, F..

The EXO-200 detector, part II: auxiliary systems

The EXO-200 experiment searched for neutrinoless double-beta decay of 136 Xe with a single-phase liquid xenon detector. It used an active mass of 110 kg of 80.6%-enriched liquid xenon in an ultra-low background time projection chamber with ionization and scintillation detection and readout. Here, this paper describes the design and performance of the various support systems necessary for detector operation, including cryogenics, xenon handling, and controls. Novel features of the system were driven by the need to protect the thin-walled detector chamber containing the liquid xenon, to achieve high chemical purity of the Xe, and to maintain thermal uniformity across the detector.

47 OTHER INSTRUMENTATION↗

Trace radioactive impurities in final construction materials for EXO-200

Here, we report results from a systematic measurement campaign conducted to identify low radioactivity materials for the construction of the EXO-200 double beta decay experiment. Partial results from this campaign have already been reported in a 2008 paper by the EXO collaboration. Here we release the remaining data, collected since 2007, to the public. The data reported were obtained using a variety of analytic techniques. The measurement sensitivities are among the best in the field. We concluded the construction of the EXO-200 detector and Phase-I data was taken from 2011 to 2014. The detector’s extremely low background implicitly verifies the measurements and the analysis assumptions made during construction and reported in this paper.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗