Engineering topics
Hayes, Christopher
Publications and source records attributed to Hayes, Christopher.
Aerosol Observing System (AOS): Two Spot Aethalometer a1-level datastream
Raw data from aethelometer 2 spot model
aosaeth (01)
The aethalometer provides a real-time readout of the concentration of “black” or “elemental” carbon aerosol particles (BC or E) in an air stream. It measures the rate of change of optical transmission through a spot on a filter where aerosol is continuously collected and calculates the concentration of optically absorbing material in the sample. With a built-in reference check of the optical source and a mass flowmeter to monitor air flow rate, the aethalometer calculates the mean BC content of the sampled air stream and provides on-the-spot readout.
aosuhsas (01)
The ultra-high-sensitivity aerosol spectrometer (UHSAS) is an optical-scattering, laser-based, aerosol particle spectrometer system for sizing particles in the 60 to 1000 nanometer (nm) range [1–3]. The instrument counts particles in up to 100 user-specified sizing bins. The instrument’s laser illuminates particles, which scatter light. The system captures the peak light signals that are generated. These signals are used for particle sizing, since the amount of light scattered correlates strongly with particle size. The UHSAS is part of the aerosol observing system (AOS).
aospsap3w1s.b1
The aospsap3w1m process reads in data from the aospsap3w.b1 datastream and applies a dilution correction to the aerosol sample flow rate, and a filter loading correction to the absorption coefficients. The filter loading is corrected by multiplying by the Weiss transfer function f_w = 1/1.031 * 1/1.22 * (1/(1.0796 Tr + 0.71)) = 1 / (1.3579 Tr + 0.8931) where Tr is the transmittance for a given wavelength. This form incorporates the initial PSAP transfer function with modifications in Bond 1999 and subsequently in Ogren 2010. The one second (1s) data stream is only created for BNL psap3w instruments where the 1 second samples are available. The one minute (1m) data streams contain the averages of the one second data.
aosccn100 (01)
The cloud condensation nuclei counter measures the concentration of aerosol particles by drawing an air sample through a column with thermodynamically unstable supersaturated water vapor that can condense onto aerosol particles. Particles that are activated, i.e., grown larger in the process, are counted and sized by an optical particle counter (OPC). In this way the CCN measures activated ambient aerosol particle number concentration as a function of supersaturation. ARM operates both one- and two-column versions of the CCN.
AOS: Humidified Tandem Differential Mobility Analyzer (a1-level)
ARM Instrument: Humidified Tandem Differential Mobility Analyzer
aosnephwet (c1)
The nephelometer measures the total scattering and hemispheric backscattering of aerosol. Nephelometers are deployed in pairs, with one measuring the ambient conditions and the other measuring the scattering as a function of slowly increasing or decreasing relative humidity (RH). The combination can then be used to derive the hygroscopic growth factor as a function of RH: f(RH). The nephelometer is part of the aerosol observing system (AOS).
aospsap3w (c1)
The particle soot absorption photometer collects aerosol particles on a substrate and measures the change in light transmission relative to a reference filter. Bulk particle absorption is derived after correcting for scattering effects. The PSAP is part of the aerosol observing system (AOS).
aosnephdry (c1)
The nephelometer measures the total scattering and hemispheric backscattering of aerosol. Nephelometers are deployed in pairs, with one measuring the ambient conditions and the other measuring the scattering as a function of slowly increasing or decreasing relative humidity (RH). The combination can then be used to derive the hygroscopic growth factor as a function of RH: f(RH). The nephelometer is part of the aerosol observing system (AOS).
AOS: ambient nephelometer measurements (b1-level)
AOS reference nephelometer, harmonized
aosaeth (a1)
The aethalometer provides a real-time readout of the concentration of “black” or “elemental” carbon aerosol particles (BC or E) in an air stream. It measures the rate of change of optical transmission through a spot on a filter where aerosol is continuously collected and calculates the concentration of optically absorbing material in the sample. With a built-in reference check of the optical source and a mass flowmeter to monitor air flow rate, the aethalometer calculates the mean BC content of the sampled air stream and provides on-the-spot readout.