Effects of temperature fluctuations of IUE data quality
Analysis of IUE calibration lamp images shows that variation in the temperature of the scientific instrument causes shifts in the location of the spectral format with respect to the reseau grid on the detector and in the location of the reseaux themselves. In high dispersion, a camera head amplifier temperature difference of 6C corresponds to a shift of 4 pixels in the spectral format for LWR and 2 pixels for SWP along the dispersion direction. Shifts perpendicular to the disperson (for the same temperature difference) are less than one pixel for both cameras. In low dispersion spectra, the shifts are similar but orthogonal to those described above with the larger motion lying in the direction perpendicular to the dispersion. In both dispersion modes, the observed shifts are apparently independent of wavelength. In high dispersion, the constant pixel shift mimics a constant velocity error.