A first test of CUPID prototypal light detectors with NTD-Ge sensors in a pulse-tube cryostat

CUPID is a next-generation bolometric experiment aiming at searching for neutrinoless double-beta decay with ~250 kg of isotopic mass of 100 Mo. It will operate at ~10 mK in a cryostat currently hosting a similar-scale bolometric array for the CUORE experiment at the Gran Sasso National Laboratory (Italy). CUPID will be based on large-volume scintillating bolometers consisting of 100 Mo-enriched Li 2 MoO 4 crystals, facing thin Ge-wafer-based bolometric light detectors. In the CUPID design, the detector structure is novel and needs to be validated. In particular, the CUORE cryostat presents a high level of mechanical vibrations due to the use of pulse tubes and the effect of vibrations on the detector performance must be investigated. In this paper we report the first test of the CUPID-design bolometric light detectors with NTD-Ge sensors in a dilution refrigerator equipped with a pulse tube in an above-ground lab. Light detectors are characterized in terms of sensitivity, energy resolution, pulse time constants, and noise power spectrum. Despite the challenging noisy environment due to pulse-tube-induced vibrations, we demonstrate that all the four tested light detectors comply with the CUPID goal in terms of intrinsic energy resolution of 100 eV RMS baseline noise. Indeed, we have measured 70–90 eV RMS for the four devices, which show an excellent reproducibility. We have also obtained high energy resolutions at the 356 keV line from a 133 Ba source, as good as Ge semiconductor γ detectors in this energy range.

47 OTHER INSTRUMENTATION↗

Twelve-crystal prototype of Li 2 MoO 4 scintillating bolometers for CUPID and CROSS experiments

An array of twelve 0.28 kg lithium molybdate (LMO) low-temperature bolometers equipped with 16 bolometric Ge light detectors, aiming at optimization of detector structure for CROSS and CUPID double-beta decay experiments, was constructed and tested in a low-background pulse-tube-based cryostat at the Canfranc underground laboratory in Spain. Performance of the scintillating bolometers was studied depending on the size of phonon NTD-Ge sensors glued to both LMO and Ge absorbers, shape of the Ge light detectors (circular vs. square, from two suppliers), in different light collection conditions (with and without reflector, with aluminum coated LMO crystal surface). The scintillating bolometer array was operated over 8 months in the low-background conditions that allowed to probe a very low, μBq/kg, level of the LMO crystals radioactive contamination by 228 Th and 226 Ra.

47 OTHER INSTRUMENTATION↗

CUPID: The Next-Generation Neutrinoless Double Beta Decay Experiment

We report CUPID is a next-generation tonne-scale bolometric neutrinoless double beta decay experiment that will probe the Majorana nature of neutrinos and discover lepton number violation in case of observation of this singular process. CUPID will be built on experience, expertise and lessons learned in CUORE and will be installed in the current CUORE infra-structure in the Gran Sasso underground laboratory. The CUPID detector technology, successfully tested in the CUPID-Mo experiment, is based on scintillating bolometers of Li 2 MoO 4 enriched in the isotope of interest 100 Mo. In order to achieve its ambitious science goals, the CUPID collaboration aims to reduce the backgrounds in the region of interest by a factor 100 with respect to CUORE. This performance will be achieved by introducing the high efficient α/β discrimination demonstrated by the CUPID-0 and CUPID-Mo experiments, and using a high transition energy double beta decay nucleus such as 100 Mo to minimize the impact of the gamma background. CUPID will consist of about 1500 hybrid heat-light detectors for a total isotope mass of 250 kg. The CUPID scientific reach is supported by a detailed and safe background model based on CUORE, CUPID-Mo and CUPID-0 results. The required performances have already been demonstrated and will be presented.

72 PHYSICS OF ELEMENTARY PARTICLES AND FIELDS↗

Optimization of the first CUPID detector module

CUPID will be a next generation experiment searching for the neutrinoless double β decay, whose discovery would establish the Majorana nature of the neutrino. Based on the experience achieved with the CUORE experiment, presently taking data at LNGS, CUPID aims to reach a background free environment by means of scintillating Li 2 100 MoO 4 crystals coupled to light detectors. Indeed, the simultaneous heat and light detection allows us to reject the dominant background of α particles, as proven by the CUPID-0 and CUPID-Mo demonstrators. In this work we present the results of the first test of the CUPID baseline module. In particular, we propose a new optimized detector structure and light sensors design to enhance the engineering and the light collection, respectively. We characterized the heat detectors, achieving an energy resolution of (5.9 ± 0.2) keV FWHM at the Q-value of 100 Mo (about 3034 keV). We studied the light collection of the baseline CUPID design with respect to an alternative configuration which features gravity-assisted light detectors’ mounting. In both cases we obtained an improvement in the light collection with respect to past measures and we validated the particle identification capability of the detector, which ensures an α particle rejection higher than 99.9%, fully satisfying the requirements for CUPID.

72 PHYSICS OF ELEMENTARY PARTICLES AND FIELDS↗

CUORE opens the door to tonne-scale cryogenics experiments

The past few decades have seen major developments in the design and operation of cryogenic particle detectors. This technology offers an extremely good energy resolution – comparable to semiconductor detectors – and a wide choice of target materials, making low temperature calorimetric detectors ideal for a variety of particle physics applications. Rare event searches have continued to require ever greater exposures, which has driven them to ever larger cryogenic detectors, with the CUORE experiment being the first to reach a tonne-scale, mK-cooled, experimental mass. CUORE, designed to search for neutrinoless double beta decay, has been operational since 2017 at a temperature of about 10 mK. This result has been attained by the use of an unprecedentedly large cryogenic infrastructure called the CUORE cryostat: conceived, designed and commissioned for this purpose. In this article the main characteristics and features of the cryogenic facility developed for the CUORE experiment are highlighted. In this work, a brief introduction of the evolution of the field and of the past cryogenic facilities are given. The motivation behind the design and development of the CUORE cryogenic facility is detailed as are the steps taken toward realization, commissioning, and operation of the CUORE cryostat. The major challenges overcome by the collaboration and the solutions implemented throughout the building of the cryogenic facility will be discussed along with the potential improvements for future facilities. The success of CUORE has opened the door to a new generation of large-scale cryogenic facilities in numerous fields of science. Broader implications of the incredible feat achieved by the CUORE collaboration on the future cryogenic facilities in various fields ranging from neutrino and dark matter experiments to quantum computing will be examined.

72 PHYSICS OF ELEMENTARY PARTICLES AND FIELDS↗

The CUORE cryostat: An infrastructure for rare event searches at millikelvin temperatures

The CUORE experiment is the world's largest bolometric experiment. The detector consists of an array of 988 TeO 2 crystals, for a total mass of 742 kg. CUORE is presently taking data at the Laboratori Nazionali del Gran Sasso, Italy, searching for the neutrinoless double beta decay of 130 Te. A large custom cryogen-free cryostat allows reaching and maintaining a base temperature of ~10 mK, required for the optimal operation of the detector. This apparatus has been designed in order to achieve a low noise environment, with minimal contribution to the radioactive background for the experiment. In this paper, we present an overview of the CUORE cryostat, together with a description of all its sub-systems, focusing on the solutions identified to satisfy the stringent requirements. We briefly illustrate the various phases of the cryostat commissioning and highlight the relevant steps and milestones achieved each time. Finally, we describe the successful cooldown of CUORE.

73 NUCLEAR PHYSICS AND RADIATION PHYSICS↗

First Results from CUORE: A Search for Lepton Number Violation via 0 ν β β Decay of Te 130

The CUORE experiment, a ton-scale cryogenic bolometer array, recently began operation at the Laboratori Nazionali del Gran Sasso in Italy. The array represents a significant advancement in this technology, and in this work we apply it for the first time to a high-sensitivity search for a lepton-numberviolating process: 130 Te neutrinoless double-beta decay. Examining a total TeO 2 exposure of 86.3 kg yr, characterized by an effective energy resolution of d7.7 ± 0.5T keV FWHM and a background in the region of interest of d0.014 ± 0.002T counts=dkeV kg yrT, we find no evidence for neutrinoless double-beta decay. Including systematic uncertainties, we place a lower limit on the decay half-life of T 0ν 1/2 ( 130 Te) > 1.3 × 10 25 yr (90% C.L.); the median statistical sensitivity of this search is 7.0 × 10 24 yr. Combining this result with those of two earlier experiments, Cuoricino and CUORE-0, we find T 0ν 1/2 ( 130 Te) > 1.5 × 10 25 yr (90% C.L.), which is the most stringent limit to date on this decay. Interpreting this result as a limit on the effective Majorana neutrino mass, we find m ββ < (110 - 520) meV, where the range reflects the nuclear matrix element estimates employed.

72 PHYSICS OF ELEMENTARY PARTICLES AND FIELDS↗

The large enriched germanium experiment for neutrinoless double beta decay (LEGEND)

We report the observation of neutrinoless double-beta decay (0νββ) would show that lepton number is violated, reveal that neu-trinos are Majorana particles, and provide information on neutrino mass. A discovery-capable experiment covering the inverted ordering region, with effective Majorana neutrino masses of 15 - 50 meV, will require a tonne-scale experiment with excellent energy resolution and extremely low backgrounds, at the level of ~0.1 count /(FWHM·t·yr) in the region of the signal. The current generation 76 Ge experiments GERDA and the Majorana Demonstrator, utilizing high purity Germanium detectors with an intrinsic energy resolution of 0.12%, have achieved the lowest backgrounds by over an order of magnitude in the 0νββ signal region of all 0νββ experiments. Building on this success, the LEGEND collaboration has been formed to pursue a tonne-scale 76 Ge experiment. Lastly, the collaboration aims to develop a phased 0νββ experimental program with discovery potential at a half-life approaching or at 10 28 years, using existing resources as appropriate to expedite physics results.

72 PHYSICS OF ELEMENTARY PARTICLES AND FIELDS↗

Limit on the radiative neutrinoless double electron capture of 36 Ar from GERDA Phase I

Neutrinoless double electron capture is a process that, if detected, would give evidence of lepton number violation and the Majorana nature of neutrinos. Here, a search for neutrinoless double electron capture of 36 Ar has been performed with germanium detectors installed in liquid argon using data from Phase I of the GERmanium Detector Array (Gerda) experiment at the Gran Sasso Laboratory of INFN, Italy. No signal was observed and an experimental lower limit on the half-life of the radiative neutrinoless double electron capture of 36 Ar was established: T 1/2 > 3.6 × 10 21 years at 90% CI.

74 ATOMIC AND MOLECULAR PHYSICS↗

CUORE-0 detector: design, construction and operation

The CUORE experiment will search for neutrinoless double-beta decay of 130 Te with an array of 988 TeO 2 bolometers arranged in 19 towers. CUORE-0, the first tower assembled according to the CUORE procedures, was built and commissioned at Laboratori Nazionali del Gran Sasso, and took data from March 2013 to March 2015. In this study we describe the design, construction and operation of the CUORE-0 experiment, with an emphasis on the improvements made over a predecessor experiment, Cuoricino. Finally, in particular, we demonstrate with CUORE-0 data that the design goals of CUORE are within reach.

46 INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND ↗

Status of the CUORE and results from the CUORE-0 neutrinoless double beta decay experiments

CUORE is a 741 kg array of TeO 2 bolometers for the search of neutrinoless double beta decay of 130 Te. The detector is being constructed at the Laboratori Nazionali del Gran Sasso, Italy, where it will start taking data in 2015. If the target background of 0.01 counts/(keV·kg·y) will be reached, in five years of data taking CUORE will have a 1σ half life sensitivity of 10 26 y. CUORE-0 is a smaller experiment constructed to test and demonstrate the performances expected for CUORE. The detector is a single tower of 52 CUORE-like bolometers that started taking data in spring 2013. The status and perspectives of CUORE will be discussed, and the first CUORE-0 data will be presented.

72 PHYSICS OF ELEMENTARY PARTICLES AND FIELDS↗

Results on $\beta \beta $ decay with emission of two neutrinos or Majorons in $^{76}$ Ge from GERDA Phase I: GERDA Collaboration

A search for neutrinoless ββ decay processes accompanied with Majoron emission has been performed using data collected during Phase I of the GERmanium Detector Array (GERDA) experiment at the Laboratori Nazionali del Gran Sasso of INFN (Italy). Processes with spectral indices n = 1,2,3,7 were searched for. No signals were found and lower limits of the order of 10 23 yr on their half-lives were derived, yielding substantially improved results compared to previous experiments with 76 Ge. A new result for the half-life of the neutrino-accompanied ββ decay of 76Ge with significantly reduced uncertainties is also given, resulting in T 2ν 1/2 = (1.926 ± 0.094) × 10 21 yr.

73 NUCLEAR PHYSICS AND RADIATION PHYSICS↗

Search for Neutrinoless Double-Beta Decay of with CUORE-0

We report the results of a search for neutrinoless double-beta decay in a 9.8 kg yr exposure of 130 Te using a bolometric detector array, CUORE-0. The characteristic detector energy resolution and background level in the region of interest are 5.1 ± 0.3 keV FWHM and 0.058 ± 0.004 (stat.) ± 0:002 (syst.) counts/(keV kg yr), respectively. The median 90% C.L. lower-limit sensitivity of the experiment is 2.9 x 10 24 yr and surpasses the sensitivity of previous searches. We find no evidence for neutrinoless double-beta decay of 130 Te and place a Bayesian lower bound on the decay half-life, T 0$_1$ 1/2 > 2.7 x 10 24 yr at 90% C.L. Combining CUORE-0 data with the 19.75 kg yr exposure of 130 Te from the Cuoricino experiment we obtain T 0$_1$ 1/2 > 4.0 x 10 24 yr at 90% C.L. (Bayesian), the most stringent limit to date on this half-life. Using a range of nuclear matrix element estimates we interpret this as a limit on the effective Majorana neutrino mass, m ββ < 270 - 760 meV.

72 PHYSICS OF ELEMENTARY PARTICLES AND FIELDS↗

CUORE-0 results and prospects for the CUORE experiment

With 741 kg of TeO{sub 2} crystals and an excellent energy resolution of 5 keV (0.2%) at the region of interest, the CUORE (Cryogenic Underground Observatory for Rare Events) experiment aims at searching for neutrinoless double beta decay of {sup 130}Te with unprecedented sensitivity. Expected to start data taking in 2015, CUORE is currently in an advanced construction phase at LNGS. CUORE projected neutrinoless double beta decay half-life sensitivity is 1.6 × 10{sup 26} y at 1σ (9.5 × 10{sup 25} y at the 90 % confidence level), in five years of live time, corresponding to an upper limit on the effective Majorana mass in the range 40–100 meV (50–130 meV). Further background rejection with auxiliary bolometric detectors could improve CUORE sensitivity and competitiveness of bolometric detectors towards a full analysis of the inverted neutrino mass hierarchy. CUORE-0 was built to test and demonstrate the performance of the upcoming CUORE experiment. It consists of a single CUORE tower (52 TeO{sub 2} bolometers of 750 g each, arranged in a 13 floor structure) constructed strictly following CUORE recipes both for materials and assembly procedures. An experiment its own, CUORE-0 is expected to reach a sensitivity to the ββ(0ν) half-life of {sup 130}Te around 3×10{sup 24} y in one year of live time. We present an update of the data, corresponding to an exposure of 18.1 kg y. An analysis of the background indicates that the CUORE performance goal is satisfied while the sensitivity goal is within reach.

72 PHYSICS OF ELEMENTARY PARTICLES AND FIELDS↗

Improvement of the energy resolution via an optimized digital signal processing in GERDA Phase I

An optimized digital shaping filter has been developed for the Gerda experiment which searches for neutrinoless double beta decay in $^{76}$Ge. The Gerda Phase I energy calibration data have been reprocessed and an average improvement of 0.3 keV in energy resolution (FWHM) corresponding to 10% at the $Q$ value for $0\nu \beta \beta $0νββ decay in $^{76}$Ge is obtained. This is possible thanks to the enhanced low-frequency noise rejection of this Zero Area Cusp (ZAC) signal shaping filter.

72 PHYSICS OF ELEMENTARY PARTICLES AND FIELDS↗

The CUORE and CUORE-0 experiments at Gran Sasso

The Cryogenic Underground Observatory for Rare Events (CUORE) is an experiment to search for neutrinoless double beta decay ($0\nu\beta\beta$) in $^{130}$Te and other rare processes. CUORE is a cryogenic detector composed of 988 TeO$_2$ bolometers for a total mass of about 741 kg. The detector is being constructed at the Laboratori Nazionali del Gran Sasso, Italy, where it will start taking data in 2015. If the target background of 0.01 counts/(keV$\cdot$kg$\cdot$y) will be reached, in five years of data taking CUORE will have an half life sensitivity around $1\times 10^{26}$ y at 90\% C.L. As a first step towards CUORE a smaller experiment CUORE-0, constructed to test and demonstrate the performances expected for CUORE, has been assembled and is running. The detector is a single tower of 52 CUORE-like bolometers that started taking data in spring 2013. The status and perspectives of CUORE will be discussed, and the first CUORE-0 data will be presented.

71 CLASSICAL AND QUANTUM MECHANICS, GENERAL PHYSIC↗

The CUORE and CUORE-0 experiments at Gran Sasso

The Cryogenic Underground Observatory for Rare Events (CUORE) is an experiment to search for neutrinoless double beta decay (0νββ) in 130 Te and other rare processes. CUORE is a cryogenic detector composed of 988 TeO 2 bolometers for a total mass of about 741 kg. The detector is being constructed at the Laboratori Nazionali del Gran Sasso, Italy, where it will start taking data in 2015. If the target background of 0.01 counts/(keV·kg·y) will be reached, in five years of data taking CUORE will have an half life sensitivity around 1 × 10 26 y at 90% C.L. As a first step towards CUORE a smaller experiment CUORE-0, constructed to test and demonstrate the performances expected for CUORE, has been assembled and is running. The detector is a single tower of 52 CUORE-like bolometers that started taking data in spring 2013. The status and perspectives of CUORE will be discussed, and the first CUORE-0 data will be presented.

73 NUCLEAR PHYSICS AND RADIATION PHYSICS↗

CUORE and beyond: Bolometric techniques to explore inverted neutrino mass hierarchy

The CUORE (Cryogenic Underground Observatory for Rare Events) experiment will search for neutrinoless double beta decay of 130 Te. With 741 kg of TeO 2 crystals and an excellent energy resolution of 5 keV (0.2%) at the region of interest, CUORE will be one of the most competitive neutrinoless double beta decay experiments on the horizon. With five years of live time, CUORE projected neutrinoless double beta decay half-life sensitivity is 1.6 × 10 26 y at 1σ (9.5 × 10 25 y at the 90% confidence level), which corresponds to an upper limit on the effective Majorana mass in the range 40–100 meV (50–130 meV). Further background rejection with auxiliary light detector can significantly improve the search sensitivity and competitiveness of bolometric detectors to fully explore the inverted neutrino mass hierarchy with 130 Te and possibly other double beta decay candidate nuclei.